Volume 3 Issue 1
Jun.  2019
Turn off MathJax
Article Contents
Rongrong Pang, Libo Zhang, Yong Wang, Wenping Han, Min Huang, Qiang Fu. Optimization and performance validation for the fully automatic blood type analysis system[J]. Blood&Genomics, 2019, 3(1): 67-72. doi: 10.46701/APJBG.2019012019112
Citation: Rongrong Pang, Libo Zhang, Yong Wang, Wenping Han, Min Huang, Qiang Fu. Optimization and performance validation for the fully automatic blood type analysis system[J]. Blood&Genomics, 2019, 3(1): 67-72. doi: 10.46701/APJBG.2019012019112

Optimization and performance validation for the fully automatic blood type analysis system

doi: 10.46701/APJBG.2019012019112
  • Publish Date: 2019-06-30
  • To evaluate and validate the application of fully automatic blood type analysis system parameters under actual lab conditions. All key system parameters were optimized and validated accordingly. The optimized parameters were centrifugal speed at 550 rpm; centrifugal time 20 min; resuspension speed 1,200 rpm; resuspension time: 45 s; incubation temperature 25℃ ; incubation time 400 s; incubation rate: 0 rpm. The sampled red blood cell concentration was 3%. The ratio of plasma to red blood cells reagent was 60 μL:30 μL; the ratio of antibody (reagent) to sampled diluted red blood cell was 30 μL:30 μL. After applying our key parameters for optimization and validation, the automatic blood type detection system's performance was found to meet the relevant requirements, effectively improving the accuracy and reliability of the detection system.

     

  • loading
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索

    Article Metrics

    Article views (742) PDF downloads(1054) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return